MAIN
MENU
CATALOG INDEX BY ALPHABETICAL ORDER
CATALOG
INDEX BY DEPARTMENT
For more info: JSi@jsits.com
- [ 1100
Series
Visible Spectrophotometer
] - [ 2100
Series Visible & UV/Vis
Spectrophotometer ] - [ 800
Series Spectrophotometers
] -[
Smart
Spectrophoto- meter Model
2000 ] - [ Spectrophotometer
] - [ UV/VIS
Spectro- photometer model V-530 ] - [
UV/Visible
spectrophotometer ] |
UV/VIS Spectrophotometer
Model V-530
The new Model V-530 Spectrometer is the most versatile,
economical and reliable instrument in the general purpose UV/Visible
class. Like the more sophisticated V-series instruments, the V-530 is
based on more than 25 years of Jasco experience with advanced optics and
state-of-the-art computer technology as well as inputs from thousands of
users worldwide. The V-530 proves a highly accurate and reliable
instrument able to fit a variety of applications from routine analysis
through advanced research.
The V-530 can be configured for individual needs offering control by
either a Windows® based PC or by the simplified intelligent
Remote Module (iRM) both via the standard RS-232C interface. The low cost
iRM offers easy data acquisition and processing using a large backlit
touch-screen LCD display designed for quick, simplified routine
applications. The V-530 can be quickly upgraded to full PC control by
connecting the RS-232C cable to the standard interface provided on all
PC’s. The new Spectra Manager is the same 32-bit Windows®
program provided with all higher end of spectroscopy products. Spectra
Manager provides the operator with a wide range of tools to enable
flexible data acquisition modes, complete data processing, custom printing
and instrument validation. Because Spectra Manager is common for all
spectroscopy products, data files from Fluorescence, FTIR, CD and other
instruments can be easily imported for comparison, display and
printing.
Features
- Simplicity and ease of use
Both the iRM and PC software
are written in an intuitive manner allowing even first time users to
measure with confidence.
- Compact design
The optical bench of the V-530 is a compact
484(w) x 477(d) x 205(h) mm saving space in the crowded modern
laboratory.
- Excellent optical performance
Double beam technology and
high-quality concave grating result in excellent spectroscopic
performance. Reducing stray light to 0.04% enables highly precise
measurement in a wide photometric range.
- High speed scanning
Bright optics and fast response
detectors allow the V-530 to scan at speeds up to 4000nm/min. without
tracking error.
- Versatility
A full complement of sampling accessories
such as cell holders, flow cells, temperature controllers and software
packages such as Kinetic and Protein-nucleic acid analysis are available
to optimize the V-530 for specific applications.
- Validation
For laboratories concerned about GMP/GLP
compliance, the V-530 offers an optional validation software package to
ensure instrument and data integrity.
Specifications
- Optical system: Double beam, single mono-chromator
with a 1200 grooves/mm concave grating and moditied Rowland mount
- Spectral bandwidth: 2nm
- Wavelength range: 190 ~ 1100 nm
- Light source: D2 lamp for UV range (190 ~ 350 nm),
Halogen lamp for VIS range (330 ~ 1100 nm)
- Light source switching: Selectable within a range
of 330 to 350 nm
- Wavelength repeatability: ±0.1 nm
- Wavelength accuracy: ±0.5 nm
- Photometric modes: ABS, %T, %R
- Photometric range: -2 ~ 3 ABS
- Photometric reproducibility:
±0.001 ABS (0 ~
0.5 ABS) ±0.002 ABS (0.5 ~ 1 ABS) ±0.15%T
- Photometric accuracy:
±0.002 ABS (0 ~ 0.5
ABS) ±0.004 ABS (0.5 ~ 1 ABS) ±0.3%T (Above figures obtained
using NIST SRM 930D)
- Stary light: 0.04%T both at 220 nm with Nal 10 g/l
solution and at 340 nm with NaNO2 50 g/l solution
- Wavelength scan speed: 4000, 2000, 1000, 400, 200,
100, 40 nm/min
- Wavelength slew speed: 8000 nm/min
- Sampling interval: 0.1, 0.2, 0.5, 1, 2 nm/data
(spectral scan mode);
0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10 sec/data (time
scan mode)
- Baseline stability: ±0.001 ABS/hour
Response:
Slow (250 nm, lamp on longer than 60 min, temperature variation ±5°C)
- Baseline flatness: ±0.001 ABS within 200 ~ 1000 nm
range after baseline correction
Response: medium Scan speed: 100
nm/min
- Response: Quick, Fast, Medium' slow
- Detector: Silicon photo-diode (S1337)
- Output port: RS-232C interface
- Power requirement: 130 W
- Dimensions: 484 W x 477 D x 205 H mm
- Net weight: Approx. 16 kg
CAT.
NO. |
DESCRIPTION |
6736-J004A |
V-530 UV/VIS Spectrophometer
(basic unit) Contents: -
Spectrophotometer (1) - AC power cable
(1) - RS-232C cable (1) - Spare
fuses (2) - Hexagonal wrench
(1) - Instruction manual (1) |
6736-J003A |
IRM-559 intelligent Remote
Module |
4880-0400C |
VWS-580 Control and data
processing software (Spectra
Manager) | |