MAIN MENU

CATALOG INDEX
BY ALPHABETICAL
ORDER

CATALOG INDEX
BY DEPARTMENT

For more info:
JSi@jsits.com

- [ 435A/B Humidity /
Temperature Sensor ]
- [
435E Relative
Humidity / Air
Temperature Sensor ]
- [
Aquaterr Digital Soil
Moisture, Temperature
and Salinity Meters ]
- [
Concrete Moisture
Meter ]
- [ Digital Wood
Moisture Meter ]
- [ External Temperature/
Humidity Sensor ]
- [ Heavy Duty Moisture
Meter ]
- [ Humidity &
Temperature Meter
- [ Irrometer Soil
Moisture Sensor and
Digital Meter ]
- [ Kelway Soil pH and
Moisture Meter ]
- [ Leaf Wetness Sensor ]
- [ Moisture Encounter ]
- [ Soil Moisture
Indicators ]
- [ Soil Moisture Tester ]
- [ Temperature Humidity
Dew Point ]
- [ Temperature Humidity
TH8 ]
- [
Testo 950 ]
- [ Wood Moisture Meter ]

  Temperature reference measuring instrument
The Testo 950

testo 400Accurate temperature measurements are becoming increasingly important, especially in regarding energy savings, indoor air quality and environmental protection.

Limitless future options available with the NEW modular system

This new modular system approach allows you to purchase the instrument that currently meets your measuring requirements, yet is capable of being upgrated at a later time for your future measuring needs.

In this way, you can add humidity and/or velocity measuring to the meter when you want. These meters are supported with our user-friendly, Basic Comfort software which can be used for programming, data processing, and even barcoding.

Our barcode pen allowes for very quick and easy data processing.

Our modular design allows for easy upgrades to the instrument as they are developed.

In this way you will always have a measuring instrument corresponding to the latest technology.

Accurate temperature measurements of liquids, semi solids, air and surfaces are performed by RTD (Pt 100) probes according to DIN IEC 751.

The instrument can be used for all measurements requiring a high accuracy e.g. measurements in chemical labs.

Description

The testo 950 measures, records and prints measurements for temperature , CO, CO2, rpm, mV and mA.

  • Numerous temperature probes for nearly any application
  • High precision - accuracy to < 0.1 °C with .01 resolution
  • Temperature probe have calibration and response adjustment capabilities

Technical data

- incl. quick extrapolation
- incl. system calibration (inst. + probe)
Pt100
Measuring Range: -200 to +800 ºC (-328 to 1472 ºF)
System Accuracy: up to ± 0.1°C / ± 0.1°F (Class A, B)
Resolution: 0.01°C (0.01°F)
NiCr-Ni
Measuring Range: -200 to +1,370°C (-328 to +2498°F)
PtRh-Pt
Measuring Range: 0 to +1,760°C (32 to +3200°F)
FeCu-Ni
Measuring Range: -200 to +1,000°C (-328 to +1832°F)
NTC
Measuring Range: -50 to +150°C (-58 to +302°F)
System Accuracy: up to 0.4°C (0.4°F)
CO measurement
Measuring Range: 0 to 500 ppm
Accuracy: ± 5 ppm
CO2 measurement
Measuring Range: 0 to 10,000 ppm
0 to 1 vol. %
Accuracy: ± 5% of m.v. or 100 ppm
rpm measurement
Measuring Range: 0.333 to 333 Hz
20 to 20,000 rpm
Accuracy: ± 1 digit
Current/voltage measurement
Measuring Range: 0 to 20 mA / 4 to 20 mA
0 to ± 10 V / 0 to ± 1 V
General technical data
Memory space
Basic version: 15 kB, cooresponds to approx.
7,500 meas. values
extended: 1 MB, corresponds to approx.
500,000 meas. values
Power supply:
alternatively:
battery / rechargeable battery
8 V mains unit (110 or 220 recharger)
Rechargeable battery lifetime: 18 h (continous oper. w.
2 thermocouple probes)
Operation temperature: 0 to +50 °C (32 to +122°F)
short-term 0 to +60°C (32 to +140°F)
Storage/transport temperature: -25 to +60°C (-13 to +140°F)
Weight: 650 g (incl. rech. battery) 1lbs. 7 ozs.
Other factores: - Automatic recognition of all connected probes
- Instrument can be upgrated (testo 650, testo 400)
- RS 232 interface for data management
Warranty: Instrument: 3 years
Probe: 1 year

There are a lot more probes available. For more information on these probes, contact us.

CAT. NO. DESCRIPTION
TE950-0000................ testo 950, 2 channel reference temperature measuring instrument (thermocouple, Pt100, NTC), incl. connection option for CO, CO2, rpm, mV/mA transducer, batteries and Li cell.