MAIN
MENU
CATALOG INDEX BY ALPHABETICAL ORDER
CATALOG
INDEX BY DEPARTMENT
For more info: JSi@jsits.com
- [ 1100
Series
Visible Spectrophotometer
] - [ 2100
Series Visible & UV/Vis
Spectrophotometer ] - [ 800
Series Spectrophotometers
] -[ Smart
Spectrophoto- meter Model
2000 ] - [ Spectrophotometer
] - [ Spectrophotometer
Model
6305 ] -
[ UV/VIS
Spectro- photometer model V-530 ] - [
UV/Visible
spectrophotometer ]
|
Spectrophotometer Model 6305
UV/Visible
- Single zenon lemp offers:
- Guaranteed 3 year life time
- Low lamp drift
- Self calibration at 3 wavelengths
- RS232 interfacing capability
- Versatile sampling system
- Small footprint
The 6350 is a high quality, low cost measurement system for
routine analysis. The 6305 features a zenon lamp which has the advantage
of offering a broad spectral source with high-energy light throughout the
entire wavelength range. In addition, with a single long lasting lamp, the
operation cost of the spectrophotometer are lowered. Uppon start up, a
three-point wavelength calibration is automatically performed at 247nm,
528nm and 824nm, assuring G.L.P. compliant operation.
The large customized LCD is easy to read, offers user-friendly
interfacing and allows simultaneous readings of wavelength and photometric
result in % Transmittance, Absorbance, or Concentration. Concentration
units can be displayed in ppm, mg/l, g/l, %, or can be left blank. The
housing is fitted with a 10mm single cell holder. In addition, a wide
variety of accessories are available, including a 4 position cell holder,
up to a 10 x 100mm cell holder, test tube holders up to 25mm in diameter,
and sipper systems. Changing the sampling configuration is quick and
simple.
Specifications
- Wavelength
- Range: 198-1000nm
- Resolution: 1nm
- Accuracy: ±2nm
- Bandwidth: 6nm
- Transmittance
- Range: 0 to 199%
- Resolution: 0.1%
- Stray light: <0.5% @ 340 & 220nm
- Accuracy: ±1%
- Absorbance
- Range: -0.300 to 1.999A
- Resolution: 0.001A
- Concentration
- Range: -3 to 1999 Concentration
- Resolution: 0.1/1
- Units: %, ppm, mg/l, g/l, M, blank
- Factor: 0 to 199.9, 1000 to 9999
- Photometric Noise: <0.001A @ 0A at 400nm
- Photmetric Stability: <0.002 AHr After 30minute warm up
- Readout: Custom LCD Graphics
- Output: Analogue (0 to 1.999V d.c.)
RS232 serial port
- Light Source: Zenon Lamp
- Input Voltage: 117/240V
- Size: 14.4 (w) x 10.7 (d) x 6.3 (h) inchces
- Weight: 13.2 lbs
CAT.
NO. |
DESCRIPTION |
635-031 |
6305 UV/Visible Spectrophotometer,
117V |
635-001 |
6305 UV/Visible Spectrophotometer,
240V |
632-001 |
Sipper Pump (240V) |
632-031 |
Sipper Pump (117V) |
634-001 |
4 Position Cell
Holder |
012-041 |
CD-AC Power
Converter |
630-020 |
Test Tube Block (13mm
diameter) |
630-021 |
Test Tube Block (25mm
diameter) |
630-022 |
Test Tube Block (16mm
diameter) |
630-005 |
20 - 100mm Cell
Holder |
630-004 |
10 x 10mm Cell
Holder |
630-028 |
Dust Cover |
035-088 |
Calibration Set |
035-079 |
100 x 10mm Glass
Cell |
035-087 |
50 x 10mm Glass Cell |
035-029 |
40 x 10mm Glass Cell |
035-086 |
20 x 10mm Glass Cell |
035-027 |
10 x 10mm Glass
Cell | |