. MAIN MENU
CATALOG INDEX BY ALPHABETICAL ORDER
CATALOG
INDEX BY DEPARTMENT
For more info: JSi@jsits.com
- [ AXSYS
Datalogger ] - [ Levelogger
Series ] - [ Pressure
Transducers: SDI Depth Sensor ] - [ Pressure
Transducers: Submersible
Depth Transmitter II ] - [ Type
A-71 Chart Recorder
] - [ Type F
Water Level Recorder ] - [
Unidata
6541 Precision Water Level
instrument
] - [ Water Level
Meter ] |
Levelogger
Series
The LT
Levelogger is unique in that the datalogger,battery, pressure transducer
and temperature sensor are all housed in a very small,minimal
maintenance,7/8"x 4.9 Ý case. The sealed design offers protection
against power surges such as nearby pumps or lightning,and greatly
simplifies maintenance. Leveloggers offer superior flexibility
compared to other water level dataloggers.They can be suspended on a
simple wireline,or connected to the surface with direct read cable for
simple downloading of data and/or reprogramming without removal from the
water.Use of the wireline further reduces costs and allows the Leveloggers
to be totally hidden from view and locked away from possible damage. A
10-year battery life,high accuracy and long-term stability all help to
make Leveloggers the ideal device for recording water levels in monitoring
and production wells,boreholes, lakes,rivers,tanks,harbours,etc.
Leveloggers are available in a variety of ranges,in the three models
described below. The fully automatic,easily programmed Levelogger
allows measurements at selected time intervals as small as 0.5 sec. with
no wraparound,so that it does not overwrite data. Logarithmic and
event-based sampling regimes are also easily programmed in the easy-to-use
software.
Applications
- Pumping and slug tests
- Watershed,drainage basin and recharge areas
- Stream gauging,lake levels and reservoirs
- Harbour and tidal fluctuation monitoring
- Wetlands and stormwater run-off monitoring
- Long term water level monitoring
- All intensive monitoring of groundwater levels
Better than Ever
Accurate,Reliable and
Easy Smaller than ever:7/8"x 4.9 Ý ((22 x 125 mm) 0.1%FS
accuracy and long-term stability Lower price Logarithmic,event-based
or linear sampling
Infra-Red Data Transfer No vented
cable or desiccant problems 2-conductor cable or simple
wireline Protected from power surges
(pumps,lightning) Maintenance-free,water-tight
Very Friendly
Software Easy data export to spreadsheets and
databases Automatic temperature compensation Barometric compensation
wizard
Telemetry Telemetry systems are available for
Leveloggers.One uses AMPs analog cellular transceivers to provide
wide-area, remote telemetry coverage,ideal for use in the USA and
Canada.Designed for use with a customer-controlled central computer
system,the software can control up to 100 remote Leveloggers,with
selectable automated reading schedules,as well as high and low level alarm
options.
LT Levelogger The LT Levelogger measures pressure and
temperature. It displays temperature readings and temperature compensated
level readings. Sampling options are linear, logarithmic or event-based
with a maximum of 2 x 24,000 readings per session. Accuracy is 0.1% FS,
from -10s¯C to 40s¯C.
Barologger The Barologger has identical specifications and
operation to the LT Levelogger in a low cost model with a low range of 5
ft. (1.5 m), suitable for monitoring barometric pressure. It allows more
accurate and easier barometric compensation than traditional vented cable
methods, without any lag time.
LTC Conductivity Levelogger LTC Leveloggers measure pressure,
temperature and conductivity. They display temperature compensated level
and conductivity readings and temperature. Sampling options are linear or
event-based with a maximum of 3 x 16,000 readings/session.
Software Leveloggers are programmed using a desktop or
portable computer and an optical reader.The software is very easy to use
and is available in DOS or Windows 95 and up.Setup and programming are
very fast and simple. The software can be used with any type of
Levelogger including previous versions of the Levelogger.There is ample
space for entering site,customer and sampling information. Leveloggers
may be synchronized to the computer clock,and there are options for
immediate start or a future start time. The battery has an estimated life
of 8-10 years and the life remaining is indicated on the programming
screen. The measurement time interval can be from 0.5 seconds to 99
hours with linear sampling.In addition,the LT Levelogger has
linear,event-based and 3 logarithmic sampling options. With the Windows
software and a Barologger,barometric compensation is highly accurate and
problem free.There is also a Barometric Compensation Wizard,which can be
used prior to export of the data to other programs. The software allows
immediate viewing of the data in graph or chart form,or simple downloading
for future examination. It also allows easy export into a spreadsheet or
database for further processing.
Levelogger Operation Programming of the Leveloggers requires
one optical reader. Place the first Levelogger into a reader attached to a
desktop or portable computer.Then fill out the various fields in the
program screen with site information and choice of sampling regime. The
Leveloggers can then be started immediatly,or with a future start
time.They can be pre-programmed and taken to the site at a later time.If
future start is chosen,no memory is used until field installation.If
immediate start is chosen the operator can see the loggers working before
deploying them. A manual measurement of the initial depth to water is
taken in each well,and noted as a base line measurement.If a Barologger is
to be used for barometric pressure measurements,it is set above high water
level in one location on site. If direct read cable has been
chosen,data can be viewed and retrieved from the Levelogger at
anytime,using the PC Interface Cable and a portable computer. To download
data from a Levelogger suspended on wireline, simply remove it from the
well and place it in an optical reader attached to a computer.This can be
done back in the office,or on site with a portable computer.Data that has
been collected is retained in the Levelogger until the Levelogger has been
reprogrammed.
LTC Conductivity Leveloggers LTC Leveloggers have a
conductivity sensor as well as the pressure and temperature sensors in the
standard LT Leveloggers.They have linear,or event-based sampling options
and a conductivity calibration wizard as well as a barometric compensation
wizard.They are ideal for use in the following applications: * Plume
monitoring and remediation sites * Salt water intrusion * Leachate
from landfills,mine tailings,waste disposal, storage sites,and more *
Agricultural and stormwater runoff * Chloride tracer tests
Use of Direct Read Cables When it is desired to get realtime
data and communicate with Leveloggers without removal from the water,they
can be deployed using direct read cables. The lower end of the direct
read cable has a miniaturized infra-red optical reader.The top cap of the
Levelogger is removed and the direct read cable attached in its place.In
turn the upper end of the cable is fitted with an infra-red connector,and
attaches,via a PC interface cable,to the portable computer. This allows
viewing of the data,downloading and/or programming in the field. The full
benefits of a sealed Levelogger with no vent tube or electrical cable
connection are also maintained.The logger is still sealed from all
electrical interference through a Faraday cage effect and cable handling
problems are minimized.
Direct Read Cable Specifications Direct read cables are
available for attachment to any Levelogger,new or old,in standard lengths
of: 50 ',100 ',200 ',300 ', 500 ' and 15m,30m,60m,80m 100m Custom
cable lengths,to fit particular monitoring situations, are also
available,as required.These cables have one marking on the cable at the
requested length.Extra cable markings,each 5 ft.or one meter,may be
requested,for situations where the direct read Leveloggers need to be used
in a variety of locations. The 1/10 "dia.(2.54 mm)cable has an HDPE
outer jacket for strength and durability.The stranded stainless steel
central conductor gives non-stretch accuracy. The upper end of the
direct read cable is fitted with a connector that can act as a well cap
for a 1 "well.This connector fits within Solinst Levelogger well caps
designed for 2 "or 4" wells,and can easily be attached at surface in other
situations.
Use of Wireline Suspension Leveloggers may also be suspended
in the water on wireline. This is a very inexpensive method of
deployment,and if in a well,allows them to be locked out of sight and
unaccessible to anyone with a special key. Solinst has adapted the
Enviro Cap ™by putting a vent hole in the cap to allow for the
equalization of barometric pressure.The well cap has a convenient eyelet
from which to suspend the Levelogger.It slips over the casing and is
locked in place with the special key,as shown. The Enviro Caps are
available sized for 2 "and 4 "wells,and well caps for other sizes of
well can also be used.
Barometric Pressure Leveloggers measure absolute pressure
(water pressure + atmospheric pressure)expressed in feet or centimeters of
water column. The most accurate method of obtaining changes in water
level is to compensate for atmospheric pressure using a Barologger.This
avoids any time lag in the compensation figures and any errors introduced
due to moisture buildup, kinking or damage to vented cable. The
Barometric Compensation Wizard in the Levelogger Windows software
simplifies the adjustment of the level measurements for barometric
pressure changes,by using the synchronized data from all Leveloggers on
site and the site Barologger. The overall results give more
reliable,highly accurate level data than that obtained when using vented
cable.
General Specifications Wetted Materials: 316 stainless steel
and Akulon Battery Life: 10 years Clock Accuracy: Better than 1
second/day Operating Temperature: -20s¯C to 80s¯C Communication:
RS232 (Optical Infra-Red) Mini LT Dimensions: 7/8" x 4.9" (22 mm x 125
mm) Mini LT Weight: 5.7 oz (160 g) LTC Dimensions: 7/8" x 10.2" (22
mm x 260 mm) LTC Weight: 10.2 oz (290 g)
Specifications |
Mini LT
Levelogger |
Barologger |
Conductivity LTC
Levelogger |
Models |
F15, F30, F60, F100,
F300, M5, M10, M20, M30, M100 |
F5/M1.5 |
F100/C5, F100/C50,
F330/C5, F330/C50 M30/C5, M30/C50, M100/C5, M100/C50 |
Memory |
Non-volatile, Flash |
Non-volatile, Flash |
|
Max. number of
readings |
2 x 24,000 Linear 2 x
19,000 Event or Log |
2 x 24,000
Linear 2 x 19,000 Event or Log |
3 x
16,000 |
Measurements
rates |
Linear 0.5sec to
99hrs 19hr, 116hr, 228 day Log or Event-Based |
Linear 0.5 sec to
99 hrs 19 hr, 116 hr, 228 day Log
Event-Based |
Linear 1sec to
99hrs Event-Based |
LevelSensor |
Ceramic
Transducer |
Ceramic
Transducer |
Ceramic
Transducer |
Normalization |
Automatic Temp
Compensation (to 1%FS from -10°C to 40°C) |
Automatic Temp
Compensation (to 1%FS from -10°C to
40°C) |
Automatic Temp
Compensation (to 1%FS from -10°C to
40°C) |
Accuracy |
0.1%FS (-10s¯C to
40s¯C) |
0.1%FS (-10°C to
40°C) |
0.1%FS (-10°C to
40°C) |
Water Level
Fluctuation |
F15/M5 = 13.12ft/4m |
|
F100/M30 =
95.12ft/29m |
Range (at sea level) |
F30/M10 =
29.52ft/9m F60/M20 = 62.32ft/19m F100/M30 =
95.12ft/29m F300/M100 = 324.72ft/99m |
|
F300/M100 =
324.72ft/99m |
Resolution |
F15/M5 =
0.007ft/0.2cm F30/M10 = 0.01ft/0.3cm F60/M20 =
0.02ft/0.6cm F100/M30 = 0.03ft/1cm F300/M100 = 0.06ft/2cm |
|
F100/M30 =
0.03ft/1cm F300/M100 = 0.06ft/2cm |
Temperature
sensor |
Spreading Resistance
Silicon |
Spreading
Resistance Silicon |
Spreading
Resistance Silicon |
Range |
-20°C to
80°C |
-20°C to
80°C |
-20°C to
80°C |
Accuracy |
0.1°C |
0.1°C |
0.1°C |
Resolution |
0.01°C |
0.01°C |
0.01°C |
Conductivity
Sensor |
N/A |
N/A |
4 Electrode
Platinum |
Normalization |
|
|
Temperature
compensation from -10°C to 40°C Specific Conductance @
25°C |
Accuracy |
|
|
1%FS |
Resolution |
|
|
0 - 5mS/cm;
1µS/cm 0 - 50mS;
10µS/cm | |