IAQ-Calc Indoor Air Quality Meters

iaq-calc_indoor_air_quality_meter.jpg (10524 octets)IAQ-Calc Meters are exceptional instruments for investigating and monitoring Indoor Air Quality (IAQ). The Model 8732 is a cost-effective meter for CO2 measurements. To monitor indoor air quality conditions, the Model 8760 and 8762 simultaneously measure and data log multiple parameters. The Model 8760 measures CO2, temperature, humidity, and calculates dew point, wet bulb temperature, absolute humidity, and % outside air. The Model 8762 adds CO.

Features and Benefits of all models

  • Stable, low-drift NDIR CO2 sensor
  • Sampling function records multiple point measurements
  • Optional portable printer provides hard copy documentation

Additional features and benefits of the model 8760 and 8762

  • Temperature and relative humidity measurements are instrumental for determining thermal comfort levels
  • Calculates % outside air from either CO2 or temperature readings
  • Direct calculation of dew point, wet bulb temperature, and absolute humidity - no psychrometric chart needed
  • Electro-chemical CO sensor for measurements made at outdoor air intakes, HVAC systems and parking garages (Model 8762 only)
  • 2-line display showing multiple parameters at same time
  • Optional detachable secondary temperature sensor

Data logging capabilities of the Model 8760 and 8762

  • Data logging ability allows user to log 14,000 samples with a time and date stamp
  • Simultaneously records all parameters available
  • Continuous data logging for full-cycle monitoring and trend analysis
  • Data can be reviewed on-screen, printed or downloaded to a computer for easy report generation
  • TSI LOGDAT downloading software permits easy transfer of data to a computer
  • LOGDAT Companion Excel Macro automatically graphs your data
  • Statistics function displays average, maximum and minimum values, and the number of recorded samples

Applications

  • Measure IAQ and comfort conditions in schools, offices and other indoor environments
  • Investigate Sick Building Syndrome
  • Comply with regulations and guidelines
  • Search for built-in-related illnesses
  • Verify building HVAC system performance
  • Examine building IAQ conditions to optimize worker productivity
Specifications
CO2 (all model)
Sensor Type Dual-channel design NDIR (non-dispersive infrared)
Range: 0 to 5000 ppm
Accuracy 3% of reading or 50 ppm, whichever is greater
Resolution 1 ppm
Response 60 seconds
Temperature (Model 8760 and 8762)
Sensor Type RTD
Range -40 to 392°F (-40 to 200°C)
Accuracy 1.0°F (0.6°C)
Resolution 0.1°F (0.1°C)
Response Time 30 seconds
Optional Secondary Temperature Probe (Model 8760 and 8762)
Sensor Type RTD
Range -40 to 392°F (-40 to 200°C)
Accuracy 1.0°F (0.6°C)
Resolution 0.1°F (0.1°C)
Response Time 20 seconds
Absolute Humidity (Model 8760 and 8762)
Range 1.51 E-05 to 0.00491 lbs moisture/ft³ (0.24 to 79.0g moisture/m³)
Accuracy Function of temperature and relative humidity
CO (Model 8762)
Sensor Type Electro-chemical
Range 0-500 ppm
Accuracy 3% of reading or 3 ppm, whichever is greater
Resolution 1 ppm
Repeatability 2% of reading
Response Time <60 sec to 90% of final value
Wet Bulb Temperature (Model 8760 and 8762)
Range 40 to 140°F (5 to 60°C)
Resolution 0.1°F (0.1°C)
Dew Point (Model 8760 and 8762)
Range 5 to 120°F (-15 to 49°C)
Resolution 0.1°F (0.1°C)
% Outside Air (Model 8760 and 8762)
Range 0 to 100%
Resolution 0.1%
Operating Temperature (all models) 41 to 113°F (5 to 45°C)
Storage Temperature (all models) -4 to 140°F (-20 to 60°C)
Logging Capability (Model 8760 and 8762)
Range Up to 14,000 samples and 1,000 test id's
Intervals 2 sec, 5 sec, 10 sec, 15 sec, 20 sec, 30 sec, 60 sec, 2 min, 5 min, 10 min, 15 min, 20 min, 30 min, 60 min.
External Meter Dimensions (all model)
Size Measurements 3.9 x 6.6 x 1.5 inches (100 x 168 xx 38 mm)
Meter Probe Dimensions (Model 8760)
Probe Length 5.81 in. (14.5 cm)
Probe Diameter 0.75 in. (1.9 cm)
Meter Probe Dimension (Model 8762)
Probe Length 6.50 (165 cm)
Probe Diameter 0.75 in (1.9 cm)
Meter Weight Dimensions (Model 8732)
Weight (with batteries) 0.76 pounds (0.34kg)
Meter Weight Dimensions (Model 8760 and 8762)
Weight (with batteries) 1.16 pounds (0.53kg)
Power (all model)
Requirements Four AA-size batteries (included) or AC adapter (optional)
Printer Interface
Type Serial
Baud Rate 1200
Model 8732 8760 8762
CO2
CO
Temperature
Humidity
% Outside Air
Dew point
Wet Bulb Temperature
Absolute Humidity
Data Logging/Downloading
Statistics
Review Data
Printer Output