MAIN
MENU
CATALOG INDEX BY ALPHABETICAL ORDER
CATALOG
INDEX BY DEPARTMENT
For more info: JSi@jsits.com
- [ IAQ-Calc
Indoor Air Quality
Meter ] - [ TH-Calc
Temperature and Humidity
Meters ]
|
IAQ-Calc Indoor Air Quality
Meters
IAQ-Calc Meters are exceptional instruments for
investigating and monitoring Indoor Air Quality (IAQ). The Model 8732 is a
cost-effective meter for CO2 measurements. To monitor indoor
air quality conditions, the Model 8760 and 8762 simultaneously measure and
data log multiple parameters. The Model 8760 measures CO2,
temperature, humidity, and calculates dew point, wet bulb temperature,
absolute humidity, and % outside air. The Model 8762 adds CO.
Features and Benefits of all
models
- Stable, low-drift NDIR CO2 sensor
- Sampling function records multiple point
measurements
- Optional portable printer provides hard copy
documentation
Additional features and benefits of the model 8760
and 8762
- Temperature and relative humidity measurements are
instrumental for determining thermal comfort levels
- Calculates % outside air from either CO2 or
temperature readings
- Direct calculation of dew point, wet bulb temperature,
and absolute humidity - no psychrometric chart needed
- Electro-chemical CO sensor for measurements made at
outdoor air intakes, HVAC systems and parking garages (Model 8762
only)
- 2-line display showing multiple parameters at same
time
- Optional detachable secondary temperature sensor
Data logging capabilities of the Model 8760 and
8762
- Data logging ability allows user to log 14,000 samples
with a time and date stamp
- Simultaneously records all parameters available
- Continuous data logging for full-cycle monitoring and
trend analysis
- Data can be reviewed on-screen, printed or downloaded
to a computer for easy report generation
- TSI LOGDAT downloading software permits easy transfer
of data to a computer
- LOGDAT Companion Excel Macro automatically graphs your
data
- Statistics function displays average, maximum and
minimum values, and the number of recorded samples
Applications
- Measure IAQ and comfort conditions in schools, offices
and other indoor environments
- Investigate Sick Building Syndrome
- Comply with regulations and guidelines
- Search for built-in-related illnesses
- Verify building HVAC system performance
- Examine building IAQ conditions to optimize worker
productivity
Specifications |
CO2
(all model) |
Sensor Type |
Dual-channel design NDIR
(non-dispersive infrared) |
Range: |
0 to 5000 ppm |
Accuracy |
3% of reading or 50 ppm,
whichever is greater |
Resolution |
1 ppm |
Response |
60 seconds |
Temperature (Model 8760 and
8762) |
Sensor Type |
RTD |
Range |
-40 to 392°F (-40 to
200°C) |
Accuracy |
1.0°F (0.6°C) |
Resolution |
0.1°F (0.1°C) |
Response Time |
30 seconds |
Optional Secondary Temperature Probe (Model
8760 and 8762) |
Sensor Type |
RTD |
Range |
-40 to 392°F (-40 to
200°C) |
Accuracy |
1.0°F (0.6°C) |
Resolution |
0.1°F (0.1°C) |
Response Time |
20 seconds |
Absolute Humidity (Model 8760 and
8762) |
Range |
1.51 E-05 to 0.00491 lbs
moisture/ft³ (0.24 to 79.0g moisture/m³) |
Accuracy |
Function of temperature
and relative humidity |
CO
(Model 8762) |
Sensor Type |
Electro-chemical |
Range |
0-500 ppm |
Accuracy |
3% of reading or 3 ppm,
whichever is greater |
Resolution |
1 ppm |
Repeatability |
2% of reading |
Response Time |
<60 sec to 90% of
final value |
Wet
Bulb Temperature (Model 8760 and 8762) |
Range |
40 to 140°F (5 to
60°C) |
Resolution |
0.1°F (0.1°C) |
Dew
Point (Model 8760 and 8762) |
Range |
5 to 120°F (-15 to
49°C) |
Resolution |
0.1°F (0.1°C) |
%
Outside Air (Model 8760 and 8762) |
Range |
0 to 100% |
Resolution |
0.1% |
Operating Temperature
(all models) |
41 to 113°F (5 to
45°C) |
Storage Temperature (all
models) |
-4 to 140°F (-20 to
60°C) |
Logging Capability (Model 8760 and
8762) |
Range |
Up to 14,000 samples and
1,000 test id's |
Intervals |
2 sec, 5 sec, 10 sec, 15
sec, 20 sec, 30 sec, 60 sec, 2 min, 5 min, 10 min, 15 min, 20 min,
30 min, 60 min. |
External Meter Dimensions (all
model) |
Size Measurements |
3.9 x 6.6 x 1.5 inches
(100 x 168 xx 38 mm) |
Meter
Probe Dimensions (Model 8760) |
Probe Length |
5.81 in. (14.5 cm) |
Probe Diameter |
0.75 in. (1.9 cm) |
Meter
Probe Dimension (Model 8762) |
Probe Length |
6.50 (165 cm) |
Probe Diameter |
0.75 in (1.9 cm) |
Meter
Weight Dimensions (Model 8732) |
Weight (with
batteries) |
0.76 pounds
(0.34kg) |
Meter
Weight Dimensions (Model 8760 and 8762) |
Weight (with
batteries) |
1.16 pounds
(0.53kg) |
Power
(all model) |
Requirements |
Four AA-size batteries
(included) or AC adapter (optional) |
Printer Interface |
Type |
Serial |
Baud Rate |
1200 |
|
|
|
|
Model |
8732 |
8760 |
8762 |
CO2 |
|
|
|
CO |
|
|
|
Temperature |
|
|
|
Humidity |
|
|
|
% Outside Air |
|
|
|
Dew point |
|
|
|
Wet Bulb Temperature |
|
|
|
Absolute Humidity |
|
|
|
Data Logging/Downloading |
|
|
|
Statistics |
|
|
|
Review Data |
|
|
|
Printer Output |
|
|
|
NIST Calibration Certificate |
|
|
| |