Oil Content Analyser
Model OCMA-350

Pushbutton simplicity, perfect for field analysis!
Compact and simple, the OCMA-350 oil content analyzer delivers quick,accurate measurements in the lab or in the field.It efficiently measures the level of oil in water or soil for environmental applications, checks for residual oil on semiconductor parts which have been cleaned, and measures oil on any industrial surface. Using the OCMA-350 is almost effortless.Simply inject the extract from your sample, dissolved in S-316 solvent, into the measuring cell, and press one button. You can even use your PC to simplify operation.

 

Recommended Applications

 

The OCMA-350 has various special features to assure accurate measurements.

Stable readout
The unit's data evaluation function analyzes the readings, determines when the value has stabilized, and displays taht stable value.
Self-monitoring
Messages on the LCD screen inform you of any elctrical malfunctions, irregularities in measurement, and deterioration of parts, so that you can always be assured of proper functioning.
Simplified recordkeeping
The OCMA-350 keeps a record of the time and date of measurement along with each data set it records. When printing out, it provides you with both the time and the measurement, allowing for easy recordkeeping.

 

SPECIFICATIONS

Application :Oil contamination of fresh and salt water;soil analysis
Principle :Solvent extraction,NDIR analysis (infrared spectroscopy)
Detector :Pyroelectric sensor
Cell :Quartz cell,cell length 20 mm
Measuring range and unit :0 to 200 mg/l,0 to 1000 mg/kg,0 to 1 Abs.
Resolution :mg/l ;0 to 99.9mg/l;0.1 mg/l 100 to 200 mg/l;1 mg/l mg/kg ;0 to 9.99 mg/kg; 0.01 mg/kg 10.0 to 99.9 mg/kg;0.1 mg/kg 100 to 1000 mg/kg;1 mg/kg Abs ;0 to 1.000 Abs; 0.001 Abs
Repeatability :mg/l ;0 to 9.9 mg/l;+0.4 mg/l +1 digit 10.0 to 99.9 mg/l;+2.0 mg/l +1 digit 100 to 200 mg/l;+4 mg/l +1 digit Abs ;+1%F.S.
Measurement :Manually controlled
Calibration :Automatic calibration after the calibration standard is introduced to the instrument
Extraction solvent : S-316 solvent
Sample/solvent volume :Approx.6.5 ml,ratio of sample to solvent is 1:1
Display :Measured value;31/2digits LCD with back-light Message;Character display LCD with back-light (16 x 2 characters)
Functions :Self diagnostics,Auto hold function,Calendar clock
Output :RS-232C and centronics printer port
Ambient temperature :0 to 40 ”ĘC
Power requirement :100 to 240 V AC ”¾10%,50/60 Hz,60 VA
Dimensions :200(H)x250(W)x285(D)mm 7.9(H)x9.8(W)x11.0(D)in
Weight :Approx.5 kg/11 lb

Model SR-300 Solvent Reclaimer (optional)
Highly efficient, the SR-300 solvent reclaimer can reduce your per test solvent cost by up to 90%. This optional unit, designed especially for reclaiming S-316 solvent, features a double column of activated carbon and activated aluminium. It has a large filtering capacity, is easy to operate, and requires no electricity.
External dimensions:
200 (W) x 200 (D) x 600 (H) mm
7.9 (W) x 7.9 (D) x 23.6 (H) in

ORDER NUMBER

MODEL NUMBER

DESCRIPTION

350751

OCMA-350

Oil concentration measuring analyzer

350752

SR-300

Solvent reclaimer