Fourier Transform Infrared Spectrometer
model FT/IR-600 series

fourier_transform_infrared_spectrometer_600series.jpg (7204 octets)The new FT/IR-600 series is the top of the line in research grade FT/IR¡¯s. Designed to complement the recently released FT/IR-400 series the 600 series utilizes the same innovative Spectra Manager software running under Windows¢ç. The FT-IR-600 features a larger and more versatile optical bench to meet the resultion, sensitivity and wavenumber range requirements demanded in research grade FT/IRs.

Features

Versatility
Three main FT/IR-600 models are available with either a purgeable or an evacuable interferometer

Optical performance
Standard on all FT/IR-600 Spectrometers:

End purging while enhancing sensitivity
Unique for this range of instruments is the availability of three relatively inexpensive models with evacuable interferometers. Vacuum type FT/IRs (Models FT/IR-610V, 615V, and 620V) are usually considered the logical choice for far infrared applications: an expensive, by effective, alternative to conventional flushing with dry air and nitrogen. However, JASCO¡¯s three relatively inexpensive vacuum models are also well suited for mid infrared applications. Removing water and CO2 absorption bands from the background allows higher sensitivity over the entire wavelength range.

Aluminum alloy casing enables vacuum operation
The aluminum alloy interferometer casing offers durability, stability and high sealing efficiency. Nitrogen purge capability is standard on all models of the 600 series while the 600V models can be efficiently evacuated by an inexpensive vacuum pump. The vacuum interferometer is more efficient at removing water vapor and CO2 than the traditional purging by nitrogen or dry gas. The vacuum offers significant improvement not only in the far IR region but also in the mid IR where measurement is effected by absorption of water vapor. For the ultimate removal of vapor and CO2 the All V models can be made fully evacuable including detector chamber and sample compartment. (Factory option)

Light sources, detectors, beam splitters
The FT/IR-600 Series can be equipped with a range of sources, beam splitters and detectors to expand the standard 7800 to 350 cm-1 range to cover 15000 to 50 cm-1. Three separate light sources can be incorporated in the unit including ceramic, Tungsten and high pressure Mercury. Up to four PC controlled detectors can be used, up to two within the main unit and two attached at the external port. Beam splitters can be exchanged by any operator and the auto-alignment feature adjusts for maximum throughput in about 15 minutes.

Rapid scanning
The rapid scan feature can acquire up to 20 scans per second allowing reaction profiles of rapidly changing compounds to be easily monitored. The kinetic change associated with an instantaneous adhesive. GC-FTIR also typically utilizes the rapid scan function to monitor GC separations. The rapid scan feature is standard on the FT/IR-620 and optional for the 610/615 models.

Vibration Isolation
All FT/IR-600 models feature a specially designed vibration insulator that completely separates the optical bench from both the upper and lower casing. This isolation provides extremely stable scanning allowing accurate measurement even during the highest sensitivity measurements.

FT/Raman Upgrade
The FT/IR-620 can be easily upgraded to measure FT/Raman by coupling the Model RFT-600 Raman attachment and a YAG laser (locally available). The Gold coated optics offer the highest throughput to provide the sensitivity demanded by Raman measurement. Both Micro and Macro FT/Raman can be measured and the sample stage allows both vertical and horizontal positioning.

Spectra Manager
The new Spectra Manager software allows control of a wide variety of instrumentation from the same common platform. Spectra Manager controls not only the FTIR series instruments but also V-series UV/VIS/NIR Spectrophotometers, J-Series CD/ORD Spectropolarimeters, and FP-750 Fluorescence Spectrofluorometer. The integrated package contains all the functions necessary to process all forms of spectroscopic data and frequently used functions can be conveniently accessed from user selectable tool bars. When measurement is completed, the analysis program is automatically started up, measurement data is transferred, and the spectrum is displayed in a view window.

Data processing / Handling

Specifications of FT/IR-610/615/620

Model

FT/IR-610

FT/IR-615

FT-IR-620

Hardware specifications

Measurable W.N. range

7800 - 350 cm-1

Expandable W.N. range

15000 - 50 cm-1

(W.N. accuracy)

(Less that ¡¾0.01cm-1 , theoritical value)

Resolution

0.5, 1, 2, 4, 8, 16 cm-1

0.25, 0.5, 1, 2, 4, 8, 16 cm-1

Optiocal system

Single beam

Aperture

Variable by software in 7 steps

Interferometer

28¡Æ incident Michelson interferometer, corner cube mirror, sealed and purgeable

Interferometer drive mechanism

Mechanical bearing, electromagnetic drive

Interferometer drive speed

0.5, 1, 2, 4, 8 mm/sec, auto; 4 mm/sec (for MCT), 2 mm/sec (for DLATGS)

0.25 mm/sec (with FT-Raman option

Rapid scan

Option

20 spectra /sec, standard

Beam splitter

Ge coated KBr as standard (CaF2, Csl, Mylar... as optional)

Light source

High temperature ceramic source as standard (W, Hg... optional)

Detector

Thermally insulated DLATGS as standard (MCTs, InSb... optional)
Up to two detector can be housed inside the main unit.

Sensitivity (S/N ratio)

25000:1, 4 cm-1 resolution, near 2200 cm-1, peak to peak, 1 min measurement

A/D converter

18 bit plus autoranging function

Dimensions

770 (W) x 610 (D) x 320 (h_ mm

Weight

Approx. 75 kg

Standard compositions
1 main unit, 1 Spectra Manager for Windows 95, 1 instruction manual, 1 acessories set pins, 1 power cable, 1 hexagonal wrench, 1 polystyrene film, 1 dust cover.

CAT. NO.

DESCRIPTION

6749-J041A

FT/IR-610 FTIR Spectrometer, 0.5 cm-1

6749-J043A

FT/IR-615 FTIR Spectrometer, 0.25 cm-1

6749-J042A

FT/IR-620 FTIR Spectrometer, 0.25 cm-1

6749-J0541A

FT/IR-610V FTIR Spectrometer, 0.5 cm-1

6749-J053A

FT/IR-615 FTIR Spectrometer, 0.25 cm-1

6749-J052A

FT/IR-620 FTIR Spectrometer, 0.25 cm-1

Optional beam splitter

6749-J201A

CaF2, 15000 - 1000 cm-1

6749-J203A

Csl, 5000 - 220 cm-1 (factory option)

6749-J2041A

Mylar (5 Fm), 600 - 150 cm-1

6749-J205A

Mylar (12 Fm), 2600 - 50 cm-1

Optional detectors

6749-J211A

DTGS(PE), 700 - 50 cm-1

6749-J212A

MCT (mid), 10000 - 650 cm-1

6749-J2131A

InSb, 15000 - 1000 cm-1

Light source

6749-J2211A

Halogen, 15000 - 2200 cm-1

6749-J213A

High-press, Hg, 100 -50 cm-1, (factory option)

IR microscope

6694-J012A

Micro-20-16 IR microscope

6694-J005A

Micro-20-38 IR microscope

Optional software

4880-0701A

Quantitative analysis

4880-0702A

JASCO library search

4880-0703A

Validation, auto-sampling

4880-0704A

Validation software, only