MAIN MENU
CATALOG INDEX BY ALPHABETICAL ORDER
CATALOG
INDEX BY DEPARTMENT
For more info: JSi@jsits.com
- [ 4
& 3 Gas Monitors - Low cost &
basic ] - [
450
Photoionization Detector ] - [ AIM
000 Series and 250 Series Gas Detectors
] - [ AIM
600 Millennium Compact Gas
Monitor ] - [ AIM
915 series "stand alone" gas detector ] - [
CO Alarm
- residential ] -
[ Pocket-size
Portable Radiation Alert,
Monitor
] - [ Gas
Data LMSxi
] - [ Q-Check
CO2 and CO Meter ] - [ Shirtpocket Size
Ozone
Sensor ] - [ Simple,
Inexpensive Ozone Monitor ] -
[ Single
Gas Portable Monitors ] - [ Testo 300 ] - [
Testo 400
] - [ Testo
450 ] - [ TIF5GA
Automovite exhaust gas
detector ] - [ TIF8800A
Combus- tible
Gas Detector ] - [ VOC Gas
sensors ] - [ VOC Monitor ] -
[ Yes-205-5 Five
Channel IAQ Datalogger ] |
Gas Data
LMSxi
Portable multi-channel environmental monitor - up to 7 gas
analysis, pressures, flow rates, and temperature measurements in one
unit.
Infra red Methane*, Carbon dioxide*, LEL* and
Hydrocarbons. Oxygen*, atmospheric pressure*, differential pressure,
flow rate and temperature.
*Standard
channels |
|
- Intrinsically safe to Eex ib IIB T2
- Modular design for easy upgrade
- 4 internal gas expansion points
- Serial interface allows upload of site, bore hole and historic
data.
- Fully programmable pump operation
- Auto peak CH4 detection
- Auto flush function
- Multifunction data storage
- Automatic unattended data logging
- Clear easy to use operator controls
General |
Ex. Rating to: |
EEx ib IIB T2 |
Ambient Temp Range |
-10 to 40°C |
Battery Life |
10 hours typical |
Aspiration Rate |
300 ml/min typical |
Storage Capacity |
>900 Readings |
Protection Rating |
IP65 |
Dimensions |
150 x 200 x 190 mm (approx) |
Weight |
<2.0 Kg |
Standard Channels |
|
Range |
Typical Accuracy |
Typical Response |
Methane |
0 to 100% |
0.2% @ 5% |
20s |
|
|
1.0% @ 30% |
|
|
|
3.0% @ 100% |
|
|
0 to 100% LEL |
4% LEL |
30s |
Carbon Dioxide |
0 to 100% |
0.1% @ 10% |
20s |
|
|
3.0% @ 50% |
|
|
|
3% @ 100% |
|
Oxygen |
0 to 25% |
0.5% |
20s |
Atmospheric Pressure |
800 to 1200 mbar |
5 mbar, 1 mbar resolution |
20s |
Standard Accessories |
Battery Charger, Carry Case, Sample Pipe & Filter,
Manual, Calibration Certificate |
Optional Gas Channels (Maximum of 4) See Note
1 |
CO2 |
10000 ppm |
5% of fs |
30s |
H2S |
200 ppm |
5% of fs |
30s |
CO |
1000 ppm |
5% of fs |
30s |
NO2 |
20 ppm |
5% of fs |
30s |
SO2 |
100 ppm |
5% of fs |
30s |
HCN |
100 ppm |
5% of fs |
30s |
Hydrocarbon Gas |
%, LEL or ppm depending on compound. |
Relative Humidity |
100% |
5% of fs |
30s |
Other Optional Channels |
|
Range |
Typical Accuracy |
Resolution |
Differential Pressure (low range) (see notes 2& 3) |
±1000 Pa |
2% of fs |
1 Pa |
Differential Pressure (high range) (see notes 2& 3) |
±100 mbar |
2% of fs |
0.1 mbar (see note 4) |
Temperature |
-10 to 60°C |
0.5°C |
0.1°C |
Flow Monitoring (see note 3) |
Bore hole Flow & Pressure Expansion (low flow) |
Extraction Flow & Pressure Expansion (high
flow) |
Flow Range |
+30/-10 l/hour |
0.1 to 1000 m2/hr |
Flow Resolution |
0.1 l/hour |
0.1 m3/hr |
Differential Pressure |
+300/-30 Pa (Typical) |
±30 mbar (see note 4) |
Static Pressure |
N/A |
±400 mbar |
Notes: 1. Many other gas types are available as expansion
channels. 2. Only one differential pressure range can be added. 3.
Differential pressure is automatically included with flow options. 4.
Internal resolution of mbar range differential pressure is 0.01 mbar.
Readings are stored and downloaded with this increased