Abbe Refractometer Model
DR-M
The Model DR-M Series Abbe Refractometers, featuring multiple
wavelengths and automatic Abbe Number measurement are the first
instruments of their kind for the high-level Refractive Index analysis of
liquids, glass, plastic, & film.
Ease to Use.
Operation consists of placing a sample on the prism; inserting an
interference filter in the light source; while looking through the
eyepiece, the control knob is turned until the shadowline is centered in
the crosshairs (see image below left); the reading is shown on the large
LCD display. Units calibrate with water (Model DE-M2) or the included
glass standard.
Professional Features.
Large, menu-driven LCD display; primary & secondary prisms; color
compensator dial; large measurement dial; focusable eyepiece; desiccant
case; drain nozzle.
Multiple Wavelengths.
Measures at 589nm (D) or 546nm (E) with the two included interference
filters. Capable of measuring in the range of 450-1100nm with the use of
optional interference filters. For testing above 680nm, the optional
Infrared Viewer (cat. no. RE-9115) is required. External halogen light
source features digital readout, variable intensity, and adjustable
positioning of the light guide.
Abbe Number Measurement.
Automatic, digital readout of Abbe Number (vD & vE) in 4 easy
steps. For vD, the included 589nm interference filter and the optional
486nm (cat. no. RE-3521) & 656nm (cat. no. RE-3522) interference
filters are used. For vE, the included 546nm interference filter and the
optional 480nm (cat. no. RE-3524) & 644nm (cat. no. RE-3525
interference filters are used.
Digital Accuracy.
High resolution and accuracy provides substantial precision for both
liquid and solids testing. The digital display and sample operation both
enhance overall precision - no analog scales to read.
Prism Temperature Control.
Featuring 2 circulator nozzles for the primary prism and 2 for the
secondary prism, the DR-M Series models can be connected to a circulating
bath to accuracy regulate prism temperature. The prism temperature is
displayed on the LCD display. The effective temperature range is
5-50°C.
Glass, Plastic, 7 Film Testing.
Transparent or translucent glass and plastic samples can be measured
with the use of the included contact liquid. Transparent or translucent
film samples can be measured with the use of the included contact liquid
and lighting glass. Note: The refractive index
(nD) of the sample must be lower than that of the contact liquid. The
included bromonaphthalene contact liquid has a refractive index (nD) value
of 1.6300; the included methylene iodide solution (Model DR-M4 only) has a
refractive index (nD) value of 1.7400.
Double Refraction (Birefringence)
Testing.
With the optional polarizing eyepiece (cat. no. RE-1145), the DR-M
Series models can test samples exhibiting double refraction. For example,
a piece of film with this caracteristic would have a machining direction
(Y axis), right-angle to machining direction (X axis), and a thinkness
direction (Z axis) -- the refractive index must be measured for each
axis.
DR-M Series
Specifications |
Measuring
System |
Optical Refraction
Critical Angle Detection |
Measuring Range |
Refractive Index
(model DR-M2)
- 1.3277-1.7379 [450nm]
- 1.3000-1.7100 [589nm]
- 1.2912-1.7011 [680nm]
- 1.2746-1.6843 [1100nm]
|
Refractive Index
(Model DR-M4)
- 1.5164-1.9164 [450nm]
- 1.4700-1.8700 [589nm]
- 1.4558-1.8557 [680nm]
- 1.4304-1.8303 [1100nm]
|
Resolution |
Refractive Index:
0.0001 |
Accuracy |
Refractive Index:
±0.0002 |
Wavelength
Range |
450-1100nm
(w/Interference Filters) |
Infrared
Viewer (Optional) |
Required for
681-1100nm |
Abbe
Number |
vD & vE
(w/Optional Interference Filters) 0.1
Resolution |
Interference Filters
(Included) |
589nm (D) &
546nm (E) |
Interference Filters
(Optional) |
486nm (F) &
656nm (C) [Required for Abbe Number vD] 480nm (F')
& 644nm (C') [Required for Abbe Number
vE] 450-1100nm [Curtom Wavelengths] |
Ambient
Temperature |
5-40°C |
Measuring
Temperature |
5.0-50.0°C |
Sample
Quantity |
>0.1ml |
Relative
Humidity |
<90% |
Altitude |
<5000m Above
Sea Level |
Calibration Liquid /
Medium |
Distilled Water
[Model DR-M2] Glass Standard, 1.5163 nD (w/Contact
Liquid) [Model DR-M2] Glass Standard, 1.6200 nD
(w/Contact Liquid) [Model DR-M4] |
Display |
LCD |
Measuring
Light Source |
Halogen
Lamp |
Prism |
Optical
Glass |
Prism
Housing |
SUS316 Stainless
Steel |
Circulator
Nozzles |
2 for primary
Prism, 2 for Secondary Prism |
Desiccant
Case |
Tablets w/ Color
Indicator |
Power
Supply |
100 to 240V
AC, 50/60 Hz |
Power
Consumption |
100V
A |
Printer
Output |
Centronics
(Parallel) |
Dimensions |
29(L)x13(W)x31(H)
cm [Main Unit] 33(L)x15(W)x11(H) cm [Light
Source] |
Weight |
6.0kg [Main
Unit] 3.0kg [Light Source] |
Supplied
With |
- Light Source (1 ea.)
- Power Card (1 ea.)
- Light Support (1 ea.)
- Light Guide (1 ea.)
- Interference filter, 589nm (1 ea.)
- Interference filter, 546nm (1 ea.)
- Glass Standard, 1.5163 nD (1 ea.) [Model DR-M2]
- Glass Standard, 1.6200 nD (1 ea.) [Model DR-M4]
- Contact Liquid, Bromonaphthalene, 4ml Bottle (1
ea.)
- Contact Liquid, Methyl Iodide Solution, 4ml Bottle (1
ea.) [Model DR-M4]
- Lighting Glass for Film Measurement (1 ea.)
- Halogen Lamp (1 ea.)
- Tubing Band (10 ea.)
- Allen Wrench (1 ea.)
- Instruction Manual (1 ea.)
|
CAT.
NO. |
DESCRIPTION |
1410 |
- Model DR-M2
- Scale: Refractive Index (nD)
- Range: 1.3000-1.7100
- Resolution: 0.0001
- Accuracy: ±0.0002
- Temperature Range: 5.0-50.0°C
- ATC: No
- Dimension: 29 x 13 x 31 cm
- Weight: 6.0kg
|
1414 |
- Model DR-M4
- Scale: Refractive Index (nD)
- Range: 1.4700-1.8700
- Resolution: 0.0001
- Accuracy: ±0.0002
- Temperature Range: 5.0-50.0°C
- ATC: No
- Dimension: 29 x 13 x 31 cm
- Weight: 6.0kg
| |