MAIN MENU

CATALOG INDEX
BY ALPHABETICAL
ORDER

CATALOG INDEX
BY DEPARTMENT

For more info:
JSi@jsits.com

- [ 3305 pH Meter ]
- [ 3310 pH Meter ]
- [ 3320 pH Meter ]
- [ Field Waterproof pH
Meter Model HI9024

and HI9025 ]
- [ HI 9110 High-tech
easy-to-use pH meter
]
- [ Kelway Soil pH and
Moisture Meter ]
- [ pH PLUS / pH meter ]
- [ Pocket pH Meter
pHep 4-5
]
- [ Portable 3 in 1 pH, TDS
and conductivity system,
model HI 9812 ]
- [ Waterproof pH/mV/°C
Meters with built-in GLP
- [ YSI 60 & 63 Field
pH Meter
]
- [ Visor pH Meter ]

3310 pH Meter

The model 3310 pH Meter is ideal for use in any busy laboratory. Ease of use and economical pricing combine to make this unit the first choice for many routine situations.

A custom liquid crystal display enables the simultaneous readout of pH (or mV) and temperature. Mode indication, prompts and error messages are presented in an easily understood format.

A simple control panel, used in conjunction with the display, ensures that even inexperienced operators can achieve reliable results. All user defined parameters are stored in the non-volatile memory and are therefore retained for immediate recall even after extended periods with power disconnected.

Calibration routines are push-button initiated and the buffer recognition facility will automatically identify and set to the appropriate temperature corrected calibration values (unless manually overridden). If an anomaly is encountered during the calibration process an explicit error message will be displayed. Sensor response can be displayed in millivolts for Redox (ORP) or ion selective measurements.

The mV reading can be made in absolute terms or the absolute value can be offset to zero to enable a relative change to be easily tracked. And end point detection icon is used to indicate when readings have reached a steady stade.

Temperature measurement is displayed and used for automatic temperature compensation to correct for the changes in electrode response with variations in temperature. A manual temperature value can be entered if desired.

The Model 3310 can download directly to a computer or printer via its bi-directional RS232 link and provides an analog output for connection to a chart recorder or data logger.

Specifications

  • Range
    • -2 to 16.00 pH
    • ±1999mV
    • -10 to 105°C
    • +14 to 220°F
  • Resolution
    • 0.01 pH
    • 1 mV
    • ±0.1°C / ±1°F
  • Accuracy
    • ±0.02 pH
    • ±1 mV
    • ±0.5°C / ±1°F
  • Temperature Compensation : Automatic 0 to 100°C
  • Auto Buffer Recognition : 4.00, 7.00, 9.22, 10.05 (manual override)
  • Calibration : User selectable 1 or 2 point
  • Outputs : Analogue 1 mV per digit Bi-directional RS232
  • Display : LCD
  • Power : power supply 9V AC
  • Size : 275 (l) x 240 (w) x 150mm (d)
  • Weight : 1.2Kgs
CAT. NO.

JE545101-0..............

DESCRIPTION

Model 3310 pH meter with plastic bodied combination electrode, swing arm electrode Holder, ATC probe, instruction manual, 4 & 7 buffer capsules